Answer:The assembly contains two internal-source filaments mounted together on one disk.
Answer:No. The 240-MS uses one filament at a time, with the second available as a spare.
Answer:The filaments are rhenium ribbons that emit electrons when heated for ionization in the mass spectrometer.
The Agilent 240-MS ion trap filament disk is an internal ionization source service assembly containing two filaments mounted on one disk. Each electrically heated rhenium ribbon produces electrons by thermionic emission for electron ionization and related internal-source operation. The instrument uses one filament at a time, with the second available when the first reaches end of life. Filament condition can affect ionization stability, sensitivity and source performance, but a loss of signal should also prompt checks of source cleanliness, electronics and operating settings. Replacement requires shutdown, safe cooling, appropriate vacuum venting and the instrument-specific ion-source maintenance procedure. Handle the assembly without touching the fragile filament ribbons or contaminating source surfaces. After servicing, restore operating vacuum and complete tune, source and mass-spectrometer performance checks. The dual-filament disk is designed for the 240-MS internal ionization assembly; it is not a general GC inlet consumable.
Unless otherwise stated, the minimum order amount is USD 1,000.
For orders below USD 1,000, an additional handling fee may apply, or we may arrange local delivery / local handover where available.
All quotations are subject to final confirmation, including product availability, lead time, batch information, shipping cost, and applicable export or handling requirements.
For quotation requests, please send an RFQ from website or contact:
Please send inquiries directly to us in the “To” field. Emails sent via BCC, mass mailing, or with unrelated recipients may be automatically filtered and will not be processed.