| Agency Method|MultiLine | EPA 8327 |
|---|---|
| Brand | ZORBAX |
| Carbon Load | 10 % |
| Endcapped | No |
| Guard Column | Yes |
| Hardware | Conventional |
| Inner Diameter (ID) | 2.1 mm |
| LC Platform|MultiLine | Low Dispersion UHPLC|UHPLC| |
| Length | 5 mm |
| Maximum Temperature|MultiLine | 40 °C at pH 6-8|90 °C at pH 1-6| |
| Molecular Weight Lower Limit | 0 Da |
| Molecular Weight Upper Limit | 3000 Da |
| Particle Size | 1.8 µm |
| Particle Type | Fully Porous |
| Phase | SB-C18 |
| Pore Size | 80 Å |
| Pressure Rating | 1200 bar |
| Separation Mode | Reversed Phase |
| Shipping Solvent | Acetonitrile/Water |
| UNSPSC Code | 41115709 |
| USP Designation | L1 |
| pH Range | 0.8-8 |
Answer:Each guard is 2.1 mm ID × 5 mm, packed with 1.8 µm StableBond C18 particles.
Answer:The specified maximum pressure is 1200 bar and the phase pH range is 0.8–8.
Answer:Three UHPLC guard cartridges are supplied per package.
Agilent ZORBAX RRHD StableBond C18 guard cartridges protect compatible reversed-phase UHPLC analytical columns from particulate contamination and strongly retained sample components. Each cartridge contains 1.8 µm fully porous SB-C18 particles in a 2.1 mm internal-diameter, 5 mm-long guard format; three guards are supplied per package. The bonded phase has an 80 Å pore structure and is designed for reversed-phase separations with a specified pH range of 0.8–8. The guard is rated to 1200 bar, permitting installation in appropriately configured low-dispersion UHPLC flow paths. Choose a guard phase and internal diameter that match the analytical column and method. Install upstream of the analytical column, control sample particulate load through appropriate pretreatment, and monitor pressure and retention behavior during routine operation. Replace the guard when pressure rises or peak shape deteriorates rather than repeatedly exposing the analytical column to contaminated samples.
Unless otherwise stated, the minimum order amount is USD 1,000.
For orders below USD 1,000, an additional handling fee may apply, or we may arrange local delivery / local handover where available.
All quotations are subject to final confirmation, including product availability, lead time, batch information, shipping cost, and applicable export or handling requirements.
For quotation requests, please send an RFQ from website or contact:
Please send inquiries directly to us in the “To” field. Emails sent via BCC, mass mailing, or with unrelated recipients may be automatically filtered and will not be processed.