| Component Type | Collector body |
|---|---|
| UNSPSC Code | 41115701 |
Answer:The component is listed for Agilent 6890, 7890 and 8890 gas chromatographs with compatible FID assemblies.
Answer:No. It is a replacement component of the collector assembly, not the entire detector or ignitor system.
Answer:No. The glow-plug assembly and ignitor cable are sold separately.
The FID collector body is a replacement mechanical component within the flame ionization detector assembly used on compatible Agilent 6890, 7890 and 8890 gas chromatographs. Positioned in the detector's ion-collection region, the collector assembly is part of the electrical path used to measure ions produced in the hydrogen–air flame. Routine detector servicing may involve inspection or replacement when deposits, damage or assembly wear compromise reliable operation. The replacement collector body should be installed with the correct insulators and fasteners using the prescribed detector service sequence, without touching critical surfaces or compromising electrical isolation. The ignitor glow-plug assembly and its cable are separate components and are not included as part of the collector body. After reassembly, verify leak-free gas delivery, flame ignition, electrical stability and representative chromatographic response.
Unless otherwise stated, the minimum order amount is USD 1,000.
For orders below USD 1,000, an additional handling fee may apply, or we may arrange local delivery / local handover where available.
All quotations are subject to final confirmation, including product availability, lead time, batch information, shipping cost, and applicable export or handling requirements.
For quotation requests, please send an RFQ from website or contact:
Please send inquiries directly to us in the “To” field. Emails sent via BCC, mass mailing, or with unrelated recipients may be automatically filtered and will not be processed.