Answer:It is used in the autosampler automatic-referencing procedure to correct transport and gripper position deviations.
Answer:Yes. The calibration tool kit includes a gripper adapter.
Answer:No. It is a mechanical positioning tool for compatible preparative autosampler modules, not a detector calibration standard.
The Agilent preparative autosampler calibration tool kit is used for automatic referencing of compatible InfinityLab autosampler transport and gripper positions after relevant service work. The kit includes the gripper adapter needed by the instrument's maintenance routine. During referencing, the sampler moves to defined tray reference positions, allowing the control system to determine small positioning corrections and store the adjusted coordinates in firmware. Accurate gripper alignment is important for reliable vial handling, transport clearance and avoidance of mechanical collisions. Run the guided procedure through the instrument maintenance software and follow the required tray and fixture placement sequence. Calibration should not be attempted with unrelated tools or arbitrary vial positions. This is a service alignment fixture, not a routine injection-volume calibration standard. Verify the exact autosampler model and instrument firmware procedure before selecting the kit for a particular maintenance job.
Unless otherwise stated, the minimum order amount is USD 1,000.
For orders below USD 1,000, an additional handling fee may apply, or we may arrange local delivery / local handover where available.
All quotations are subject to final confirmation, including product availability, lead time, batch information, shipping cost, and applicable export or handling requirements.
For quotation requests, please send an RFQ from website or contact:
Please send inquiries directly to us in the “To” field. Emails sent via BCC, mass mailing, or with unrelated recipients may be automatically filtered and will not be processed.