Hach, 5440000, CL17 Chlorine Analyzer (Instrument Only)
Catalog Number: 5440000
Send an RFQ.
Category
| Accessories | Installation Kit Maintenance Kit |
|---|---|
| Accuracy | ± 5 % or ±0.03 mg/L (ppm) as CL₂, whichever is greater. *For permanganate accuracy, please refer to permanganate application notes |
| Air Purge | 0.1 cfm (0.17 m³/h) instrument quality air at max. 20 psig (ca. 1.4 bar) with 1/4-inch OD tube, quick disconnect fitting |
| Alarm | Two alarms selectable for sample concentration alarm, analyzer system warning, or analyzer system shut-down alarm. Each is equipped with an SPDT relay with contacts rated for 5A resistive load at 230 V AC. |
| Certifications | Europe, CE Approved with: EN 61326-1 CISPR 11 EN 61010-1 IEC 60529 North America: UL 61010A-1 CAN/CSA C22.2 No. 1010.1-92 |
| Communication Capabilities | 4-20mA Output |
| Cycle Time | 2.5 min |
| Description | CL17 Chlorine Analyzer (Instrument Only) |
| Display | LCD, 3-1/2 inch digit measurement readout and six-character alphanumeric scrolling text line |
| Drain | 1/2-inch ID flexible hose, hose barb |
| Enclosure Construction | ABS plastic, two clear polycarbonate windows, IP62-rated with the gasketed door latched. |
| Includes |
Unless otherwise stated, the minimum order amount is USD 1,000.
For orders below USD 1,000, an additional handling fee may apply, or we may arrange local delivery / local handover where available.
All quotations are subject to final confirmation, including product availability, lead time, batch information, shipping cost, and applicable export or handling requirements.
For quotation requests, please send an RFQ from website or contact:
Please send inquiries directly to us in the “To” field. Emails sent via BCC, mass mailing, or with unrelated recipients may be automatically filtered and will not be processed.
| CL17 Chlorine Analyser, Maintenance kit, Installation Kit, and Manual. (The power cord is ordered separately.) |
| Inlet | 1/4-inch OD polyethylene tube, quick-disconnect fitting |
|---|
| Inlet Pressure | 1 - 5 psig (0.07 to 0.34 bar), 1.5 psi is optimum |
|---|
| Interferences | Other oxidizing agents such as bromide, chlorine dioxide, permanganate and ozone will cause a positive interference. Hexavalent chromium will cause a positive interference: 1 mg/L Cr⁶⁺ = approximately 0.02 mg/L as Cl₂. Hardness must not exceed 1,000 mg/L as CaCO₃. |
|---|
| Light Source | Class 1 LED (light emitting diode) with a peak wavelength of 520 nm; 50,000 hours estimated minimum life |
|---|
| Lower Limit of Detection (LOD) | 0.03 mg/L (ppm) |
|---|
| Mounting | Wall mount |
|---|
| Operating Humidity | Up to 90 % at 40 °C (104 °F) maximum |
|---|
| Operating Temperature Range | 5 - 40 °C (41 - 104 °F) |
|---|
| Parameter | Chlorine, Free & Total |
|---|
| Pressure Range | Inlet Pressure to Sample Conditioning: 1.5 - 75 psi (0.1 - 5.2 bar) |
|---|
| Pressure Regulator Included? | No |
|---|
| Range | 0 - 5 mg/L Cl₂ |
|---|
| Recorder Outputs | One 0/4-20 mA with an output span programmable over any portion of the 0 to 5 mg/L range. Recommended load impedance 3.6 to 500 ohms, 130 V isolation from earth ground. |
|---|
| Repeatability | 0.01 ppm or 3.0 % , whichever is greater |
|---|
| Sample Flow Rate | 200 - 500 mL per minute minimum |
|---|
| Sample Temperature | 5 - 40 °C (41 - 104 °F) |
|---|
| Warranty | 12 months |
|---|
| Weight | 10.49 kg |
|---|
| What's included? | CL17 Chlorine Analyser, Maintenance kit, Installation Kit, and Manual. (The power cord is ordered separately.) |
|---|