Hach, 5500.AMC.2.KTO, 5500sc Ammonia Monochloramine Analyzer, 2 Channel
Catalog Number: 5500.AMC.2.KTO
Send an RFQ.
Category
| Accuracy | ± 5 % or 0.01 ppm (as N) for sample temperature 5 - 40 °C (41 - 104 °F); ±10% or 0.02 ppm for sample temperature 40 - 50 °C (104 - 122 °F), whichever is greater |
|---|---|
| Alarm | Four; not powered SPDT relays, each rated at 5A resistive, 240 VAC maximum |
| BABA | Eligible |
| Certifications | CE (EN 61326-1: 2006; EN 61010-1: 2010; EN 60529: 1991, +A1:2000 cETLus (UL 61010-1:2012; NEMA 250:2003; CSA C22.2 No 61010-1:2012) Australian RCM Marking |
| Communication | Modbus RS485, Profibus DP (available with external controller), 4-20mA Output |
| Cycle Time | 4.5 min |
| Enclosure Rating | IP56 / NEMA 4X |
| External Filter | No |
| Grab Sample | Grab Sample In and Grab Sample Out capability |
| Inlet | 6mm OD push-to-connect fitting for plastic tubing |
| Light Source | Class 1M LED (light emitting diode) with a peak wavelength of 650 nm |
| Lower Limit of Detection (LOD) | 0.01 ppm (as N) |
| Material | Polycarbonate (PC)/ABS case, PC Door, PC hinges and latches, 316 Stainless Steel hardware |
| Measuring Interval | 4.5 minutes |
| Mounting | Wall, panel, or table |
Unless otherwise stated, the minimum order amount is USD 1,000.
For orders below USD 1,000, an additional handling fee may apply, or we may arrange local delivery / local handover where available.
All quotations are subject to final confirmation, including product availability, lead time, batch information, shipping cost, and applicable export or handling requirements.
For quotation requests, please send an RFQ from website or contact:
Please send inquiries directly to us in the “To” field. Emails sent via BCC, mass mailing, or with unrelated recipients may be automatically filtered and will not be processed.
| Number of Channels |
|---|
| 2 |
| Number of sample streams | 1 or 2 in programmable sequence |
|---|
| Operating Humidity | 5 - 95 % non-condensing |
|---|
| Operating Temperature Range | 5 - 45 °C (41 - 113 °F) |
|---|
| Options | Standard |
|---|
| Outlet | 11mm (7/16 in.) ID slip-on fitting |
|---|
| Outputs | Four 0/4-20 mA Outputs, load impedance 600 ohm maximum |
|---|
| Parameter | Total Ammonia, Monochloramine, and Free Ammonia |
|---|
| Pollution Degree | 2/II |
|---|
| Power Requirements (Hz) | 50 - 60 Hz |
|---|
| Power Requirements (Voltage) | 100 - 240 VAC |
|---|
| Range | 0.01 - 2 ppm (as N), 0.1 - 10.0 mg/L (as Cl₂) |
|---|
| Reagent Consumption | 1L of reagents every month with 5 minute cycle time |
|---|
| Repeatability | 3 % or 0.01 ppm (as N), whichever is greater |
|---|
| Response Time | Within 5 minutes |
|---|
| Sample Flow Rate | 100 - 1000 mL/min |
|---|
| Sample Pressure | 2.5 - 87 psi (0.17 - 6.8 bar) |
|---|
| Sample Temperature | 5 - 50 °C (41 - 122 °F) |
|---|
| Storage Conditions | -20 - 60 °C (-4 - 140 °F) |
|---|
| Warranty | 12 months |
|---|
| Weight | 20.5 kg (45.19 lb) without reagents and standards, 30 kg (66.14 lb) with reagents, standards and cleaning solution |
|---|
| What's included? | 5500sc Ammonia Monchloramine Analyzer, 2 Channel with installation kit and user manuals, reagents and standards for one month use |
|---|