Hach, EZ1037.9700140T, EZ1037sc Sulfide Colorimetric Analyzer, 1 stream, 4x mA Output
Catalog Number: EZ1037.9700140T
Send an RFQ.
Category
| Ambient Temperature | 50 - 86°F (10 - 30 °C) ± 4 °C deviation at 5 - 95% relative humidity (non-condensing) |
|---|---|
| Analog Outputs | Active 0 - 20 mA (or 4 - 20 mA) max. 500 Ohm load, standard: 4, optional: 8 |
| Automatic cleaning | Yes; Frequency freely programmable: 6 hours, 12 hours, daily, weekly |
| Calibration | Automatic; 2-point; offset or slope; Frequency freely programmable: 6 hours, 12 hours, daily, weekly. Note: manufacturer recommends that a calibration is done when the reagents are replaced |
| Certifications | CE, ETL certified to UL and CSA safety standards, UKCA |
| Cycle Time | No Dilution Default: 10 min. Continuous: 9 min. Internal Dilution Default: 15 min. Continuous 12 min. |
| Demineralized Water | Internal dilution option: Rinse water & Dilution water |
| Digital Outputs | Profibus DP or Modbus RTU |
| Dimensions (H x W x D) | 688 mm x 460 mm x 340 mm |
| Drain | Atmospheric pressure, vented, min. Ø 32 mm |
| Earth Connection | Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm² |
| Instrument Air |
Unless otherwise stated, the minimum order amount is USD 1,000.
For orders below USD 1,000, an additional handling fee may apply, or we may arrange local delivery / local handover where available.
All quotations are subject to final confirmation, including product availability, lead time, batch information, shipping cost, and applicable export or handling requirements.
For quotation requests, please send an RFQ from website or contact:
Please send inquiries directly to us in the “To” field. Emails sent via BCC, mass mailing, or with unrelated recipients may be automatically filtered and will not be processed.
| Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air. Used to flush the instrument in corrosive environement. Min. 0.2 bar - Max. 0.5 bar |
| Interferences | Strong reducing agents like hydrosulfide (S2O42–), nitrite (NO2–) > 0.2 mg/L, sulfite (SO32–) >200 mg/L and thiosulfate (S2O32–) > 5 mg/L decrease the blue color and interfere. Oxygen reacts with sulfide and must be absent. Large quantities of color and turbidity interfere. Fats, oil, proteins, surfactants and tar interfere. |
|---|
| Lower Limit of Detection (LOD) | No Dilution 0.025 - 0.5 mg/L S²⁻: 0.025 mg/L Internal Dilution 0.125 - 2.5 mg/L S²⁻: 0.125 mg/L 0.25 - 5 mg/L S²⁻: 0.25 mg/L 0.625 - 12.5 mg/L S²⁻: 0.625 mg/L 1.25 - 25 mg/L S²⁻: 1.25 mg/L 1.875 - 37.5 mg/L S²⁻: 1.875 mg/L 2.5 - 50 mg/L S²⁻: 2.5 mg/L |
|---|
| Material | Hinged part: Thermoform ABS, door: PMMA Wall section: Galvanized steel, powder coated |
|---|
| Measurement method | N,N-dimethyl-p-phenylenediamine colorimetric method |
|---|
| Measurement Method | Colorimetric |
|---|
| Number of sample streams | Standard: 1 Optional accessory for 2, 4, or 8 sample streams |
|---|
| Outputs | Optional: Claros Ethernet connection and Modbus TCP/IP Ethernet connector; LAN version; 10/100 Mbps or Profinet or Ethernet IP |
|---|
| Parameter | Sulfide |
|---|
| Power | 100 - 240 VAC, 50/60 Hz Max. power consumption: 120 VA |
|---|
| Precision | Better than 2% full scale range for standard test solutions |
|---|
| Protection Class | IP44 |
|---|
| Range | No Dilution 0.025 - 0.5 mg/L S²⁻ Internal Dilution 0.125 - 2.5 mg/L S²⁻ 0.25 - 5 mg/L S²⁻ 0.625 - 12.5 mg/L S²⁻ 1.25 - 25 mg/L S²⁻ 1.875 - 37.5 mg/L S²⁻ 2.5 - 50 mg/L S²⁻ |
|---|
| Reagent Requirements | Keep between 50 - 86 °F (10 - 30 °C) |
|---|
| Relays | 5 contacts, not user configurable: Malfuntion, maintenance, analysis ready, sample ready, sample ready (EZ9150) |
|---|
| Sample Flow Rate | 100 - 300 mL/min |
|---|
| Sample Pressure | By external overflow vessel (open to atmospheric pressure) |
|---|
| Sample Quality | Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU |
|---|
| Sample Temperature | 50 - 86 °F (10 - 30 °C) |
|---|
| Validation | Automatic; Frequency freely programmable: 6 hours, 12 hours, daily, weekly |
|---|
| Warranty | 12 months |
|---|
| Weight | Max. 77 lb (35 kg) |
|---|
| What's included? | EZ1037sc Sulfide Analyzer, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets and 2 x empty 2.5L Reagent Glass Containers with Fittings (for Color, Buffer) |
|---|