Hach, EZ3508.9700140T, EZ3508sc Fluoride Ion-Selective Electrode (ISE) Standard Addition Analyzer, 1 stream, 4x mA
Catalog Number: EZ3508.9700140T
Send an RFQ.
Category
| Ambient Temperature | 50 - 86 °F (10 - 30 °C) ± 4 °C deviation at 5 - 95% relative humidity (non-condensing) |
|---|---|
| Analog Outputs | Active 0 - 20 mA (or 4 - 20 mA) max. 500 Ohm load, standard 4, optional: 8 |
| Automatic cleaning | Yes; Frequency freely programmable: 6 hours, 12 hours, daily, weekly |
| Calibration | Automatic; 2-point; offset or slope; Frequency freely programmable: 6 hours, 12 hours, daily, weekly. Note: manufacturer recommends that a calibration is done when the reagents are replaced |
| Certifications | CE, ETL certified to UL and CSA safety standards, UKCA |
| Cycle Time | No dilution Default: 5 minutes Continuous: 5 minutes Internal dilution Default: 10 minutes Continuous: 8 minutes |
| Demineralized Water | Internal dilution option: Rinse water and Dilution water |
| Digital Outputs | Relays: 5 contacts, not user configurable: Malfuntion, maintenance, analysis ready, sample ready, sample ready (EZ9150) Ethernet Connections: Optional: Claros Ethernet connection and Modbus TCP/IP Ethernet connector; LAN version; 10/100 Mbps or Profinet or Ethernet IP RS485 communication: Profibus DP or Modbus RTU |
| Dimensions (H x W x D) | 688 mm x 460 mm x 340 mm |
Unless otherwise stated, the minimum order amount is USD 1,000.
For orders below USD 1,000, an additional handling fee may apply, or we may arrange local delivery / local handover where available.
All quotations are subject to final confirmation, including product availability, lead time, batch information, shipping cost, and applicable export or handling requirements.
For quotation requests, please send an RFQ from website or contact:
Please send inquiries directly to us in the “To” field. Emails sent via BCC, mass mailing, or with unrelated recipients may be automatically filtered and will not be processed.
| Drain |
|---|
| Atmospheric pressure, vented, min. Ø 32 mm |
| Earth Connection | Earth connection Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm² |
|---|
| Instrument Air | Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air. Used to flush the instrument in corrosive environement. Min. 0.2 bar - Max. 0.5 bar |
|---|
| Interferences | Metal ions like aluminium [(Al)³+] > 72 mg/L, calcium [(Ca)²+] > 108 mg/L and iron [(Fe)²+/(Fe)³+] > 150 mg/L. Fats, oil, proteins, surfactants and tar. |
|---|
| Lower Limit of Detection (LOD) | No dilution 0.2 - 10 mg/L F - : 0.2 mg/L 0.5 - 25 mg/L F - : 0.5 mg/L 1 - 50 mg/L F - : 1 mg/L 2 - 100 mg/L F - : 2 mg/L Internal dilution 10 - 500 mg/L F - : 10 mg/L 20 - 1000 mg/L F - : 20 mg/L |
|---|
| Material | Hinged part: Thermoform ABS, Door: PMMA Wall section: Galvanized steel, powder coated |
|---|
| Measurement method | Discontinuous measurement by combined Ion-Selective Electrode with standard addition |
|---|
| Measurement Method | Ion-Selective Electrode (standard addition) |
|---|
| Number of sample streams | Standard: 1 Optional accessory for 2, 4, or 8 sample streams |
|---|
| Parameter | Fluoride |
|---|
| Power | 100 - 240 VAC, 50/60 Hz Max. power consumption: 120 VA |
|---|
| Precision | Better than 2% full scale range for standard test solutions |
|---|
| Protection Class | IP44 |
|---|
| Range | No dilution 0.2 - 10 mg/L F - 0.5 - 25 mg/L F - 1 - 50 mg/L F - 2 - 100 mg/L F - Internal dilution 10 - 500 mg/L F - 20 - 1000 mg/L F - |
|---|
| Reagent Requirements | Keep between 50 - 86 °F (10 - 30 °C) |
|---|
| Sample Flow Rate | 100 - 300 mL/min |
|---|
| Sample Pressure | By external overflow vessel (open to atmospheric pressure) |
|---|
| Sample Quality | Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU |
|---|
| Sample Temperature | 50 - 86 °F (10 - 30 °C) |
|---|
| Validation | Automatic; Frequency freely programmable: 6 hours, 12 hours, daily, weekly |
|---|
| Warranty | 12 months |
|---|
| Weight | Max. 77 lb (35 kg) |
|---|
| What's included? | EZ3508sc Fluoride Ion-Selective Electrode (ISE) Standard Addition Analyzer, 1 stream, 4x mA, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 1 x ISE Electrode, 1 x empty 5L Reagent Container with Fittings (for F Standard Solution) and 1 x empty 10L Reagent Container with Fittings (for Buffer Solution) |
|---|