Category
| Accuracy | 2% or ±10 ppb TFC, whichever is greater at pH <7.5 |
|---|---|
| Accuracy 2 | 2% or ±10 ppb HOCl, whichever is greater at pH <8 |
| Calibration Interval | 2 months |
| Calibration Method | Zero Calibration: electrically (automatic) or with de-chlorinated water Calibration: Comparison of lab method with process sample pH Calibration: Single- or two-point using standards or comparison of lab method with process sample |
| Description | 9184sc Total Free Chlorine Amperometric Sensor |
| Enclosure Rating | IP66 / NEMA 4X |
| Flow | 14 L/h (200 - 250 mL/min) auto-regulated by flow through cell |
| Interferences | No interference from chloramines. Chlorine dioxide and ozone interfere with measurement. |
| Lower Limit of Detection (LOD) | 0.005 mg/L HOCl |
| Maintenance Interval | Measurement cell: 6 months for membrane and electrolyte for typical application pH: 1 - 1.5 years for typical application |
| Measuring Interval | Continuous |
| Measuring Principle | Amperometric/Membrane (electrode, membrane, electrolyte) |
| Mounting | Flat, vertical surface (panel, stand, etc.) |
| Operating Temperature Range |
Unless otherwise stated, the minimum order amount is USD 1,000.
For orders below USD 1,000, an additional handling fee may apply, or we may arrange local delivery / local handover where available.
All quotations are subject to final confirmation, including product availability, lead time, batch information, shipping cost, and applicable export or handling requirements.
For quotation requests, please send an RFQ from website or contact:
Please send inquiries directly to us in the “To” field. Emails sent via BCC, mass mailing, or with unrelated recipients may be automatically filtered and will not be processed.
| 0 - 45 °C |
| Parameter | Total Free Chlorine: Sum of hypochlorous acid, hypochlorite and free chlorine |
|---|
| pH Range | 4 - 8 pH, acidification unit available for pH >8 |
|---|
| Pressure Range | 0.1 - 2 bar in flow cell |
|---|
| Range | 0 - 20 mg/L HOCl |
|---|
| Relative Humidity | 0 - 90 % non-condensing |
|---|
| Response Time | 90 % in less than 90 seconds |
|---|
| Sample Input | 0.25 " outer diameter |
|---|
| Sample Output | 0.5 " inner diameter |
|---|
| Sample Temperature | 2 - 45 °C |
|---|
| Standard deviation | 0.7 |
|---|
| Storage Conditions | -20 - 60 °C |
|---|
| Temperature Compensation | Automatic over sample temperature range |
|---|
| Temperature Compensation | Automatic over sample temperature range |
|---|
| Warranty | 12 months |
|---|
| Weight | 6.5 kg |
|---|
| What's included? | 9184sc Chlorine Sensor Panel with pH Sensor, Membranes and Electrolyte and Manual. |
|---|