Hach, LXV45A.99.13022, CLF10sc Free Chlorine Analyzer (Panel Only) with pHD Differential Sensor
Catalog Number: LXV45A.99.13022
Send an RFQ.
Category
| Accuracy | Free Chlorine: ± 3 % of the reference test* (DPD) at constant pH less than 7.2 (±0.2 pH unit) |
|---|---|
| Accuracy 2 | Free Chlorine: ± 10 % of the reference test** (DPD) at stable pH less than 8.5 (±0.5 pH unit from the pH at calibration) |
| Accuracy 3 | Total Chlorine: ± 10 % of the reference test** (DPD) at stable pH less than 8.5 (±0.5 pH unit from the pH at calibration) |
| Accuracy 4 | Total Chlorine: ± 20 % of the reference test** (DPD) at stable pH greater than 8.5 |
| Cable Connection | 5 pin, M12 connector |
| Cable Length | 1 m (between gateway and sc controller) |
| Calibration Method | 1-point or 2-point (zero and slope) calibration |
| Certifications | CE compliant for conducted and radiated emissions: - CISPR 11 (Class A limits) - EMC Immunity EN 61326-1 (Industrial limits) Additional certifications when connected to an sc controller |
| Connection Drain Line | 1/2 in ID |
| Connections | 1/4 in OD |
| Controller | Panel only |
| Controller Compatibility | sc controller platform |
| Description | CLF10sc Free Chlorine Analyzer (Panel Only) with pHD Differential Sensor |
Unless otherwise stated, the minimum order amount is USD 1,000.
For orders below USD 1,000, an additional handling fee may apply, or we may arrange local delivery / local handover where available.
All quotations are subject to final confirmation, including product availability, lead time, batch information, shipping cost, and applicable export or handling requirements.
For quotation requests, please send an RFQ from website or contact:
Please send inquiries directly to us in the “To” field. Emails sent via BCC, mass mailing, or with unrelated recipients may be automatically filtered and will not be processed.
| Dimensions |
|---|
| Sensor Only: 195 |mm x 25 mm (7.68 in x 0.98 in) |
| Fitting Type | USA |
|---|
| Includes | CL10sc Panel, 1 m digital extension cable, Panel Manual, pHD Sensor Manual, Chlorine Sensor Manual. |
|---|
| Interferences | Free chlorine: monochloramine, chlorine dioxide, ozone, and chalk deposits. Total chlorine: chlorine dioxide, ozone and chalk deposits |
|---|
| Length | Sensor: 195 mm |
|---|
| Lower Limit of Detection (LOD) | 30 ppb (0.03 ppm) |
|---|
| Material | Corrosion-resistant materials, fully-submersible (stainless steel, PEEK, PVC, silicon rubber and polycarbonate) |
|---|
| Measurement Method | Reagentless, electrochemical, three-electrode amperometric system |
|---|
| Mounting | Panel or Wall Mounting |
|---|
| Operating Temperature Range | 0 - 45 °C |
|---|
| Options | pHD Differential Sensor |
|---|
| Parameter | Free Chlorine |
|---|
| pH Monitoring Required? | With differential pH sensor |
|---|
| pH Range | 4 - 9 |
|---|
| Power Requirements (Voltage) | None |
|---|
| Pressure Range | 0.5 bar, no pressure impulses and/or vibrations |
|---|
| Range | 0 - 20 ppm Cl₂ |
|---|
| Repeatability | 30 ppb or 3%, whichever is greater |
|---|
| Response Time | Total Chlorine: 100 s or less for 90% change (T90) at a stable temperature and pH |
|---|
| Response Time | Free Chlorine: 140 s or less for 90% change (T90) at a stable temperature and pH |
|---|
| Sample Flow Rate | 30 - 50 L/h (7.9 - 13.2 gal/hour), Optimal is 40 L/hour (10.5 gal/hour) |
|---|
| Sample Temperature | 5 - 45°C (41 - 113°F) |
|---|
| Sampling Time | Continuous |
|---|
| Storage Conditions | -20 °C to 60 °C dry |
|---|
| Warranty | 12 months |
|---|
| Weight | Approximately 5.5 kg (12 lbs) Panel and empty panel-mounted components only |
|---|
| What's included? | CL10sc Panel, 1 m digital extension cable, Panel User Manual pHD electrode, PPS, for CL10sc (9181500), pHD Sensor User Manual Digital extension cable, 1m (6122400) Replacement Sensor for CLF10sc, SS Tip (8626200), Chlorine Sensor User Manual |
|---|