Hach, LXV525.99A12661, SC4500 Controller, Prognosys, Profibus DP, 2 Analog UPW Conductivity Sensors, 100-240 VAC, without power cord
Catalog Number: LXV525.99A12661
Send an RFQ.
Category
| Altitude | 3000 m (9842 ft) maximum |
|---|---|
| Analog Output Functional Mode | Linear, PID |
| Communication (optional) | Analog: Five 0-20 mA or 4-20 mA analog outputs on each analog output module Up to two analog Input modules (0-20 mA or 4-20 mA). Each input module replaces a digital sensor input. Digital: Profibus DPV1 module Modbus TCP Profinet IO module Ethernet IP module |
| Compatible Instruments | Compatible Sensors and Analyzers / Software Version (Release Year) Amtax sc / V2.30 (2018) or higher A-ISE sc / V1.02 or higher AN-ISE sc / V1.08 (2013) or higher N-ISE sc / V1.02 or higher Nitratax clear sc, Nitratax eco sc, Nitratax plus sc / V3.13 (2013) or higher NT3100sc/NT3200sc Phosphax sc / V2.30 (2018) or higher Phosphax sc LR/MR/HR / V1.01 (2018) or higher TSS sc / V41.73 (2013) or higher Solitax sc / V2.20 (2013) or higher TU5300sc, TU5400sc / V1.34 (2017) or higher SS7 sc (in Bypass) / V1.01 (2006) or higher Ultraturb sc / V3.06 (2017) or higher 1720E / V2.10 (2006) or higher Sonatax sc / V1.15 (2016) or higher CL17sc / V2.7 (2019) or higher CL10sc / V1.14 (2013) or higher 9184sc, 9185sc, 9187sc* / V2.03 (2013) or higher Uvas plus sc / V3.01 (2017) or higher LDO 2 sc* / V1.22 (2013) or higher 3798sc* / V2.03 (2013) or higher 3700sc + Inductive Conductive Digital Gateway 6120800 / V3.00 (2017) or higher 3422sc, Analog 3400 + Contacting Cond. Digital Gateway 6120700 / V3.00 or higher pHD sc*, pHD-S sc / V3.10 (2016) or higher 1200-S sc* / V2.04 (2013) or higher pHD analog + Digital Gateway 6120500 / V3.00 (2017) or higher RC and PC analog sensor + Digital Gateway for conventional analog pH and ORP sensors 6120600 / V3.00 (2017) or higher 8362sc* / V3.00 (2017) or higher *Hardware Version1 of instrument is not supported |
Unless otherwise stated, the minimum order amount is USD 1,000.
For orders below USD 1,000, an additional handling fee may apply, or we may arrange local delivery / local handover where available.
All quotations are subject to final confirmation, including product availability, lead time, batch information, shipping cost, and applicable export or handling requirements.
For quotation requests, please send an RFQ from website or contact:
Please send inquiries directly to us in the “To” field. Emails sent via BCC, mass mailing, or with unrelated recipients may be automatically filtered and will not be processed.
| Compatible Network Technologies | GSM 3G/4G (e.g. AT&T, T-Mobile, Rogers, Vodafone etc.) CDMA (e.g. Verizon) |
|---|
| Compliance Certifications | CE. ETL certified to UL and CSA safety standards (with all sensor types), FCC, ISED, KC, RCM, EAC, UKCA, SABS, C م (Morocco) |
|---|
| Conduit Openings | ½" NPT conduit |
|---|
| Description | Microprocessor-controlled and menu-driven controller that operates the sensor |
|---|
| Dimensions | ½ DIN - 144 x 144 x 192 mm (5.7 x 5.7 x 7.6 in.) |
|---|
| Display | 3.5-inch TFT colour display with capacitive touchpad |
|---|
| Enclosure Rating | UL50E type 4X, IEC/EN 60529–IP 66, NEMA 250 type 4X Metal enclosure with a corrosion-resistant finish |
|---|
| Indoor/Outdoor | Outdoor installation in direct sunlight or UV radiation requires UV protection screen and/or sunroof |
|---|
| Installation Category | Category II |
|---|
| Material Enclosures | Polycarbonate, aluminum (powder coated), stainless steel |
|---|
| Measurements | Two analog device connectors |
|---|
| Mounting | Wall, Pole, or Panel Mounting |
|---|
| Network Connectivity | LAN: Two Ethernet connectors (10/100 Mbps) Cellular: External 4G Wi-Fi |
|---|
| Operating Temperature Range | -20 to 60 °C (-4 to 140 °F) (8 W (AC)/9 W (DC) sensor load) -20 to 45 °C (-4 to 113 °F) (28 W (AC)/20 W (DC) sensor load) Linear derating between 45 and 60 °C (-1.33 W/°C) |
|---|
| Outputs | Profibus DP |
|---|
| Pollution Degree | 4 |
|---|
| Power Requirements (Voltage) | 100 - 240 VAC ±10%, 50/60 Hz; 1 A |
|---|
| Power Supply | No Plug |
|---|
| Protection Class | I, connected to protective earth |
|---|
| Relays | Two relays (SPDT); Wire gauge: 0.75 to 1.5 mm² (18 to 16 AWG) AC controller Maximum switching voltage: 100 - 240 VAC Maximum switching current: 5 A Resistive/1 A Pilot Duty Maximum switching power: 1200 VA Resistive/360 VA Pilot Duty DC controller Maximum switching voltage: 30 VAC or 42 VDC Maximum switching current: 4 A Resistive/1 A Pilot Duty Maximum switching power: 125 W Resistive/28 W Pilot Duty |
|---|
| Sensor Input #1 | Analog UPW Conductivity |
|---|
| Sensor Input #2 | Analog UPW Conductivity |
|---|
| Software available | Non-Claros + Embedded Prognosys |
|---|
| Storage Conditions | -20 - 70 °C (-4 - 158 °F), 0 - 95% relative humidity, non-condensing |
|---|
| USB Port | Used for data download and software upload. The controller records approximately 20,000 data points for each connected sensor. |
|---|
| Warranty | 12 months |
|---|
| Weight | 3.7 lb (controller only, w/o modules) |
|---|
| What's included? | Controller, Non-Claros + Embedded Prognosys, without power cord; includes mounting hardware |
|---|