Send an RFQ for Mettler-Toledo 30129734, DO Sensor InPro6860i/12/120/mA Ex. Iright supports global sourcing and international shipment for life science and laboratory products.
| Measurement Parameter | Oxygen |
|---|---|
| Measurement Technology | Optical fluorescence quenching |
| Accuracy | ± (1 % of the reading + 8 ppb) |
| Response Time | T98 = < 90 s |
| Primary Media/Application | Pharmaceutical/Life Science Production |
| ISM (Digital) or Analog | ISM (Digital) |
| Anti-Bubble Design | No |
| Wetted Metallic Material | 316L stainless steel |
| Surface Roughnes of Wetted Material | N5/Ra16 (Ra=0.4μm/16μin) |
| Sensor Diameter | 12 mm |
| O-Ring Material | EPDM |
| Power | 25 V DC 0.1 A |
| Process Connection | Pg 13.5 |
| Sensor Length | 120 mm |
| Wetted membrane material | PTFE |
| Connector Design | Straight |
| Model Family | InPro 6860i |
| Autoclavable/Sterilizable | Yes |
| CIP resistant | Yes |
| Measuring Temp. Range | 41 °F – 140 °F (5 °C – 60 °C) |
| Mechanical Temp. Range | -4 °F – 284 °F (-20 °C – 140 °C) |
| Operating Pressure (Bar) | 0.2–6 Bar |
| Operating Pressure (PSI) | 0–87 PSI |
Answer:It measures dissolved oxygen by optical fluorescence quenching.
Answer:This configuration has a 12 mm shaft diameter and 120 mm sensor length.
Answer:Yes. It supports qualified autoclave and sterilization cycles up to 140 °C; process integration and approvals must match the application.
The Mettler-Toledo InPro 6860i is an optical dissolved oxygen sensor for monitoring oxygen in biopharmaceutical processing and suitable bioreactor installations. It uses fluorescence quenching rather than a conventional electrochemical oxygen measurement cell. The specified sensor has a 12 mm diameter, 120 mm shaft length and a Pg 13.5 process connection. Digital communication and an mA output option support integration with compatible transmitters and process control systems. Calibration data can be retained in the intelligent sensor, assisting startup when the host transmitter supports this feature. A polished 316L stainless-steel construction and protective PTFE-coated optical sensing element support hygienic operation. The sensor is designed to tolerate qualified sterilization and autoclave cycles up to 140 °C, although the permitted process measurement conditions differ from sterilization conditions. Check installation depth, connector format, transmitter compatibility and hazardous-area approvals before commissioning. Calibrate and verify oxygen response using the validated process procedure; this sensor is not a stand-alone oxygen display instrument.
Unless otherwise stated, the minimum order amount is USD 1,000.
For orders below USD 1,000, an additional handling fee may apply, or we may arrange local delivery / local handover where available.
All quotations are subject to final confirmation, including product availability, lead time, batch information, shipping cost, and applicable export or handling requirements.
For quotation requests, please send an RFQ from website or contact:
Please send inquiries directly to us in the “To” field. Emails sent via BCC, mass mailing, or with unrelated recipients may be automatically filtered and will not be processed.
| Design Pressure |
|---|
| 87 psi (6 bar) |
| Analog Integration | 4/20 mA HART active output |
|---|
| Digital Integration | RS485 - ISM® and MODBUS RTU |
|---|
| Cable Connector | VP6 / VP8 (analog / digital) |
|---|
| Hygienic/Biocompatibility Approvals | FDA/USP Class 7 |
|---|
| Hazardous Area Approvals | ATEX/IECEx: II 1/2G Ex ia/ib HC T6 Ga/Gb II1/2D Ex ia/ib nie T83 °C Da/Db |
|---|
| Certificates & Approvals | Quality certificate, Material certificate 3.1, Surface finish certificate, ATEX/IECEx certificate, FDA/USP Class IV |
|---|
| Measurement Range | 0...saturation |
|---|
| Sterilizable/Autoclavable | Yes |
|---|
| Measured Pressure Resistance | 0 psi – 87 psi (0 bar – 6 bar) |
|---|