| Sodium Measurement Range | 0.01-100,000 ppb or equivalent ppm, auto-ranging |
|---|---|
| Sodium Measurement Accuracy | ± 10% of reading ± 0.1 ppb, typical using DIPA as reagent ± 10% of reading ± 1 ppb, typical using ammonium hydroxide as reagent |
| Response Time | 90% at 5 minutes |
| Sample pH | 2.5-12 |
| Update Rate | Once per second |
| Reagent Consumption | 3-4 L per year DIPA, or 9-12 L per year ammonium hydroxide typical more at higher temperatures and for cation exchange samples |
| Sample Flow rate | >40 mL/min (>20 mL/min for cation exchange samples), excess to drain |
| Sample Temperature | 5-50°C (41-122°F) |
| Sample Pressure | 0.3-7 bar (5-100 psig) |
| Grab Sample Measurement | Available option |
| Power Supply | 100-240 VAC, 50-60 Hz, 25 W, on power loss all settings are retained without batteries |
| Dimensions (HxWxD) | 851 mm x 400 mm x 165 mm |
| Weight | 4.5 kg (10 lbs) |
| Ambient Operating Temperature | 5-50 °C (41-122 °F) |
| Humidity | 10-90% non-condensing |
| Ratings/Approvals | CE, UL, IP65 |
Unless otherwise stated, the minimum order amount is USD 1,000.
For orders below USD 1,000, an additional handling fee may apply, or we may arrange local delivery / local handover where available.
All quotations are subject to final confirmation, including product availability, lead time, batch information, shipping cost, and applicable export or handling requirements.
For quotation requests, please send an RFQ from website or contact:
Please send inquiries directly to us in the “To” field. Emails sent via BCC, mass mailing, or with unrelated recipients may be automatically filtered and will not be processed.
| Process Connection |
|---|
| Sample inlet: 6 mm or 1/4" OD tube SS compression fitting Drain hose: 19 x 25.4 mm (¾ x 1"), 2 m (6 ft) length included |
| Analog Output | 22mA alarm, 500 ohm max load, not for use with externally powered circuit For sodium, conditioned pH, temperature four powered 0/4-20 mA, |
|---|
| Power Supply (AC/DC) | 100 V – 240 V |
|---|
| Analog Output Scaling | Linear, bi-linear, logarithmic (1,2,3,4 decades), auto ranging |
|---|
| Flow Rate | 40 mL/min – 60 mL/min |
|---|
| Measurement Parameter | Sodium |
|---|
| Electrode Conditioning | Manual |
|---|
| Operating Range (Temp.) | 5 °C – 50 °C |
|---|
| Relay Contacts | 3 A resistive freely assignable to set-points for sodium, pH, temperature |
|---|
| Sodium Calibration | Manual 3-point known addition manual 1- or 2-point |
|---|
| Operating Range | 0.01 ppb – 100,000 ppb |
|---|
| Analog Output Accuracy | ±0.05 mA |
|---|
| EnclosureType | Panel |
|---|
| Outputs | 4 Analog |
|---|
| Built-in sequencer | No |
|---|