Send an RFQ for Hamilton 6606801-01, ZEUS X1 Automated Air Displacement Pipette, PN: 6606801-01. Iright supports global sourcing and international shipment for life science and laboratory products.
| Accuracy | <1% at 1000 µL, <5% at 1 µL |
|---|---|
| Precision | <0.75% at 1000 µL, <4% at 1 µL |
| Weight | 1075 g |
| Power Requirements | 36.9-50 VDC, 2 A |
| Communication Interface | CAN (Hamilton protocol), FCC connector |
| Baud Rate | 125,500 kBaud (others on request) |
| Operating Temp High | 32 °C (90 °F) |
| Operating Temp Low | 18 °C (64 °F) |
| Operating Humidity | 15% to 85% (no condensation, indoors) |
| Storage Temp High | 70 °C (158 °F) |
| Storage Temp Low | -25 °C (-13 °F) |
| Storage Humidity | 15% to 85% (no condensation, indoors) |
| z-Drive Excursion | 200 mm |
| z-Drive Resolution | 0.1 mm |
| z-Drive Maximum Speed | 350 mm/sec |
| Volume Range | 1 - 1,000 µL |
| Pipetting Resolution | 0.1 µL |
| Maximum Pipetting Speed | 1,428 µL/s (peak speed) |
| Pipetting Mode | Single and aliquot dispensing |
| Tip Detection | Yes |
| Liquid Level Detection | Pressure-based (pLLD) and capacitive detection (cLLD) |
Unless otherwise stated, the minimum order amount is USD 1,000.
For orders below USD 1,000, an additional handling fee may apply, or we may arrange local delivery / local handover where available.
All quotations are subject to final confirmation, including product availability, lead time, batch information, shipping cost, and applicable export or handling requirements.
For quotation requests, please send an RFQ from website or contact:
Please send inquiries directly to us in the “To” field. Emails sent via BCC, mass mailing, or with unrelated recipients may be automatically filtered and will not be processed.
| Pipetting Security |
|---|
| Qualitative Pipette Monitoring (QPM) Anti-Droplet Control (ADC) |
| Operating Noise | < 60 dBA, Indoor use only |
|---|
| Device Addressing | up to 31 modules on a single CAN bus |
|---|
| Dimensions (D x H x W) | 161 mm x 335 mm x 16 mm |
|---|
| Compatible CO-RE®II Tips | 10 µL, 50 µL, 300 µL, 1 000 µL |
|---|
| RoHS Compliant | Yes |
|---|