| Chloride Measurement Range | 0-300 ppb |
|---|---|
| Chloride Limit of Detection | 0.5 ppb |
| Chloride Measurement Accuracy | ±5% of reading or ± 0.5 ppb whichever is greater |
| Sulfate Measurement Range | 0-300 ppb |
| Sulfate Limit of Detection | 1 ppb |
| Sulfate Measurement Accuracy | ±10% of reading or ±1 ppb whichever is greater |
| Measurement Cycle Time | 45 min typical, programmable between 15 minutes and 1 hour |
| Sample Flow rate | 25-50 mL/min |
| Sample Temperature | 10-45°C (50-113°F) |
| Sample Pressure | 0.3-7 bar (5-100 psig) |
| Grab Sample Measurement | Included 100 mL capacity |
| Power Supply | 100-240 VAC, 50-60 Hz, 100 W typical |
| Dimensions (HxWxD) | 927x508x305 mm (36.5"x20.9"x12") |
| Weight | 44 kg (97 lbs) |
| Ambient Operating Temperature | 10-35°C (50-95°F) |
| Humidity | 10-70% non-condensing |
| Ratings/Approvals | CE / cULus |
| Process Connection | Sample Inlet: 1/4" or 6 mm OD tube SS compressin fitting Drain Hose: 19x25.4 mm (3/4 x 1") 2 m (6 ft) length included |
Unless otherwise stated, the minimum order amount is USD 1,000.
For orders below USD 1,000, an additional handling fee may apply, or we may arrange local delivery / local handover where available.
All quotations are subject to final confirmation, including product availability, lead time, batch information, shipping cost, and applicable export or handling requirements.
For quotation requests, please send an RFQ from website or contact:
Please send inquiries directly to us in the “To” field. Emails sent via BCC, mass mailing, or with unrelated recipients may be automatically filtered and will not be processed.
| Operator Interface |
|---|
| Color touchscreen Simultaneous display of ion concentrations and analyzer status |
| Analog Output | 22mA alarm, 500 ohm max load, not for use with externally powered circuit |
|---|
| Analog Output Scaling | Linear, bi-linear, logarithmic (1,2,3,4 decades), auto ranging |
|---|
| Measurement Parameter | Chloride and sulfate |
|---|
| Relay Contacts | Mechanical rated at 250 VAC, 3 Amps (Relay 1 NC, Relay 2 to 4 NO), 4-SPDT Type Reed 250 VAC or DC, 0.5 Amps (Relay 5 to 8) |
|---|
| Analog Output Accuracy | ±0.05 mA |
|---|
| Model Family | 3000CS |
|---|
| EnclosureType | Full |
|---|
| Built-in sequencer | No |
|---|