| Application | Concentration; Diafiltration; Protein; Vaccine; Antibody |
|---|---|
| Application Area | Process development; Preclinical and clinical studies; Pilot; Production |
| Connections | DIN |
|---|---|
| Connector(s) | 0.75" Tri-clamp sanitary flange |
| Dimensions WxDxH | 120 × 190 × 225 mm |
|---|---|
| Feed / Retentate | 10 mm |
| Permeate | 10 mm |
| Additional Product Specifications | Nuts for tie rods: Bronze |
|---|---|
| Cassette Capacity | Sartocon® Cassettes: ≤ 5 |
| Cassette Format | Slice Cassettes |
| Electro Polished | Electro Polished |
| Mirror Inverted | No |
| Surface Roughness of Wetted Parts | Ra <= 0.6 µm |
| Filter- and Clamping Plate Material | 316L (1.4404/1.4435) stainless steel |
|---|
Answer:The holder accommodates up to five compatible Sartocon Slice cassettes.
Answer:No. The Sartocon Slice cassettes must be selected and ordered separately.
Answer:The stainless-steel holder supports autoclaving and in-line steam sterilization when operated within the validated equipment procedure.
The Sartocon Slice cassette holder is a reusable stainless-steel frame designed to clamp compatible Sartocon Slice cassettes for tangential flow filtration. The vertical holder arrangement is intended for low hold-up volume, venting and drainage in development and small-scale processing. It accommodates up to five Sartocon Slice cassettes, enabling an appropriate membrane area to be installed for the required application. The holder is not itself an ultrafiltration membrane: the cassette type, membrane chemistry and molecular-weight cutoff or pore size are selected and ordered separately. Assemble the unit with the specified feed, retentate and permeate connections, confirm gasket and cassette placement, and conduct pre-use integrity or leak testing as required by the process. Stainless-steel wetted components support cleaning, autoclaving and in-line steam sterilization within validated operating procedures.
Unless otherwise stated, the minimum order amount is USD 1,000.
For orders below USD 1,000, an additional handling fee may apply, or we may arrange local delivery / local handover where available.
All quotations are subject to final confirmation, including product availability, lead time, batch information, shipping cost, and applicable export or handling requirements.
For quotation requests, please send an RFQ from website or contact:
Please send inquiries directly to us in the “To” field. Emails sent via BCC, mass mailing, or with unrelated recipients may be automatically filtered and will not be processed.
| Brand | Sartocon® Slice |
|---|---|
| Filter Type | TFF |
| Sterilization Method | Autoclavable & In-Line Steam Sterilizable |
|---|
| Volume | 200 mL to 100 L |
|---|