| Accuracy | ±15% or 30 ppb, whichever is greater |
|---|---|
| Accuracy (Temperature) | ±0.5°C |
| Accuracy (mV) | ± (0.5 mV + 0.1%) |
| Alarm Output | 3 |
| Amperage | 5A |
| Analog Outputs | 2, Galvanically Isolated |
| Calibration | Double Known Addition |
| Case Material | Valox 364 |
| Certifications/Compliance | CE, CSA, UL, FCC Class A limits |
| Channel 1 | Ammonia |
| Channel 2 | Optional ph/ORP or Conductivity |
| Channels | 1 |
| Communication Details | Modbus with Additional Module |
| Current Loops | 2 |
| Description | 2110XP Ammonia Analyzer |
| Dimensions (L x W x H) | 11 x 17 x 26 in. (27 x 45 x 65cm) |
| Display Type | Custom backlit LCD with temperature, concentration, mV, error codes and menu driven prompts |
| Enclosure Rating | Waterproof; IP66; NEMA 4X |
| Flow Rate | 40mL/min. |
| Frequency | 0 to 60 Hz |
| Height (English) | 26 in. |
| Height (Metric) | 65 cm |
Answer:The analyzer measures ammonia from 0 to 10 ppm.
Answer:The analyzer is configured for one sample stream.
Answer:It provides 0–20 mA and 4–20 mA outputs for process integration.
Orion™ 2110XP Ammonia Analyzer is used for continuous online measurement and control of ammonia in boiler-water and related high-purity water processes. The system uses ammonia electrode technology to provide direct measurement across a 0–10 ppm range and delivers real-time information for optimizing chemical feed. The analyzer supports one sample stream, accepts sample temperatures from approximately 5–45°C and provides 0–20 mA or 4–20 mA outputs for process-control integration. Response performance reaches 95% within about two minutes after injection of a standard solution. The wall-mount instrument includes calibration guidance and is designed to minimize drift and reduce the need for frequent manual calibration. Its selective ammonia measurement avoids reliance on indirect pH or conductivity changes when precise boiler chemistry control is required.
Unless otherwise stated, the minimum order amount is USD 1,000.
For orders below USD 1,000, an additional handling fee may apply, or we may arrange local delivery / local handover where available.
All quotations are subject to final confirmation, including product availability, lead time, batch information, shipping cost, and applicable export or handling requirements.
For quotation requests, please send an RFQ from website or contact:
Please send inquiries directly to us in the “To” field. Emails sent via BCC, mass mailing, or with unrelated recipients may be automatically filtered and will not be processed.
| Includes |
|---|
| 2110XP ammonia analyzer with Ammonia electrode, reference electrode (100056), ATC probe (2100TP),ammonia calibration standards kit, CD user guide and options kit |
| Inlet Pressure | 8 to 100psig |
|---|
| Inputs | 85-132; 170-264 VAC |
|---|
| Keypad Type | Membrane |
|---|
| Length (English) | 11 in. |
|---|
| Length (Metric) | 27 cm |
|---|
| Max. Relative Humidity | 5 to 95%, non-condensing |
|---|
| Measurement Ranges | 0 to 10ppm |
|---|
| Mounting Options | Wall mount |
|---|
| Outputs | 0 to 20 mA; 4 to 20 mA |
|---|
| Range (mV) | ±1999.9m |
|---|
| Relay Details | 3 Relay Outputs; Relay Load: 250VAC; 30VDC; 5A |
|---|
| Resolution (Temperature) | 0.1°C |
|---|
| Resolution (mV) | 0.1mV |
|---|
| Response Time | 95% within 2 minutes of injecting a standard solution |
|---|
| Sample Inlet | 1/4 in. NPTF tube fitting |
|---|
| Sample Streams | 1 |
|---|
| Sample Temperature | 5 to 45°C |
|---|
| Voltage | 85/132 VAC, 170/264 VAC |
|---|
| Width (English) | 17 in. |
|---|
| Width (Metric) | 45 cm |
|---|
| Temperature (Metric) Operating | 5°C to 35°C |
|---|
| Type | Ammonia Analyzer |
|---|
| Unit Size | Each |
|---|