Description
Calibrated solution validation test kit. (For China: Hazardous product will be shipped from another warehouse. The corresponding invoice will be provided by Agilent China Shanghai Legal entity). Solution validation test kit comprising 12 sealed cells for completing OQ/PV of UV-Vis spectrophotometers (wavelength accuracy, photometric accuracy, stray light & resolution). Includes potassium dichromate standards of 40 mg/L, 60 mg/L, 120 mg/L & 600 mg/L plus perchloric acid blank for photometric accuracy tests; holmium perchlorate 4% in 10% perchloric acid for wavelength accuracy tests; potassium chloride solution 12 g/L, sodium nitrite solution 50 g/L & sodium iodide 1% solution plus water blank for stray light tests; 0.02% toluene in hexane plus a hexane blank for resolution tests. Each solution is sealed in a quartz cuvette. A certificate of traceability and performance is included for each solution.