| UNSPSC Code | 41115709 |
|---|---|
| Web Name Suffix Override | L x I.D. 250 mm x 4 mm, LiChroCART™ HPLC cartridge, manuCART™ holder required for use |
| Web Name Override | Superspher® 60 RP-select B (4 µm) HPLC Columns |
| NACRES | SB.52 |
| material | stainless steel column |
| agency | suitable for USP L7 |
| product line | Superspher® |
| feature | endcapped: no |
| parameter | 250 bar pressure |
| technique(s) | HPLC: suitable |
| column L × I.D. | 25 cm × 4 mm |
| matrix | fully porous particle |
| matrix active group | C8 bonding phase |
| particle size | 4 μm |
| pore size | 60 Å pore size |
| operating pH range | 2-7.5 |
| separation technique | reversed phase |
| storage temp. | no temp limit |
| Quality Level | 100 |
Answer:The LiChroCART cartridge is 250 × 4 mm with 4 µm, 60 Å C8 reversed-phase silica.
Answer:Yes. A compatible manuCART holder is required for the LiChroCART cartridge format.
Answer:No. Both may use C8 chemistry, but selectivity and method performance must be evaluated before substitution.
Supelco Superspher 60 RP-select B LiChroCART is a 250 × 4 mm analytical HPLC cartridge packed with 4 µm, 60 Å reversed-phase silica bearing C8 chemistry. The RP-select B designation identifies a specific chromatographic selectivity that should be preserved when reproducing an established method. The cartridge format requires a compatible manuCART holder and must not be treated as a stand-alone conventional screw-ended HPLC column. When moving an analytical procedure to this packing, assess retention and peak shape using the relevant test mixture and system suitability requirements. Ensure the cartridge is seated correctly in the holder before pressurizing the LC flow path. Condition the installed cartridge with a compatible mobile phase and avoid uncontrolled shifts in pH or solvent composition. Reversed-phase selectivity depends on analyte chemistry and mobile-phase formulation, so identical dimensions do not guarantee equivalence to other C8 packings. This package contains one chromatographic cartridge for use with the designated holder.
Unless otherwise stated, the minimum order amount is USD 1,000.
For orders below USD 1,000, an additional handling fee may apply, or we may arrange local delivery / local handover where available.
All quotations are subject to final confirmation, including product availability, lead time, batch information, shipping cost, and applicable export or handling requirements.
For quotation requests, please send an RFQ from website or contact:
Please send inquiries directly to us in the “To” field. Emails sent via BCC, mass mailing, or with unrelated recipients may be automatically filtered and will not be processed.