| Accuracy | ±5% or 0.01 ppb, whichever is greater |
|---|---|
| Accuracy (Temperature) | ±0.5°C |
| Accuracy (mV) | ± (0.5 mV + 0.1%) |
| Alarm Output | 3 |
| Amperage | 5A |
| Analog Outputs | 2, Galvanically Isolated |
| Calibration | Double Known Addition |
| Case Material | Valox 364 |
| Certifications/Compliance | CE, CSA, UL, FCC Class A limits |
| Channel 1 | Sodium |
| Channel 2 | Optional ph/ORP or Conductivity |
| Channels | 1 |
| Communication Details | Modbus with Additional Module |
| Current Loops | 2 |
| Description | 2111LL Low Level Sodium Analyzer |
| Dimensions (L x W x H) | 11 x 17 x 26 in. (27 x 45 x 65cm) |
| Display Type | Custom backlit LCD with temperature, concentration, mV, error codes and menu driven prompts |
| Enclosure Rating | Waterproof; IP66; NEMA 4X |
| Flow Rate | 40mL/min. |
| Frequency | 0 to 60 Hz |
| Height (English) | 26 in. |
| Height (Metric) | 65 cm |
| Includes | Instrument Only Package |
Answer:The measurement range is 0.001 ppb to 10 ppm sodium.
Answer:It is designed for continuous low-level sodium monitoring in high-purity water and industrial process streams.
Answer:No. This configuration is the instrument-only package.
Orion™ 2111LL is an online low-level sodium analyzer for continuous monitoring of trace sodium in high-purity water and related process streams. The instrument measures from 0.001 ppb to 10 ppm and is designed for applications such as steam-cycle and power-generation water monitoring where small increases in sodium can indicate contamination or process leakage. The system uses ROSS™ electrode technology, a pumpless passive-diffusion reagent system and a proprietary flow cell to support stable low-level measurements and rapid response. The instrument-only configuration provides one sodium measurement channel, backlit display, isolated analog outputs and relay alarms in a waterproof IP66/NEMA 4X enclosure. Normal sample flow is approximately 40 mL/min.
Unless otherwise stated, the minimum order amount is USD 1,000.
For orders below USD 1,000, an additional handling fee may apply, or we may arrange local delivery / local handover where available.
All quotations are subject to final confirmation, including product availability, lead time, batch information, shipping cost, and applicable export or handling requirements.
For quotation requests, please send an RFQ from website or contact:
Please send inquiries directly to us in the “To” field. Emails sent via BCC, mass mailing, or with unrelated recipients may be automatically filtered and will not be processed.
| Inlet Pressure |
|---|
| 8 to 100psig |
| Keypad Type | Membrane |
|---|
| Length (English) | 11 in. |
|---|
| Length (Metric) | 27 cm |
|---|
| Max. Relative Humidity | 5 to 95%, non-condensing |
|---|
| Measurement Ranges | 0 to 10ppm |
|---|
| Mounting Options | Wall mount |
|---|
| Outputs | 0 to 20 mA; 4 to 20 mA |
|---|
| Range | 0.001ppb to 10ppm |
|---|
| Relay Details | 3 Relay Outputs; Relay Load: 250VAC; 30VDC; 5A |
|---|
| Resolution | 1; 2; 3; or 4 significant digits |
|---|
| Response Time | 95% within 2 minutes of injecting a standard solution |
|---|
| Sample Inlet | 1/4 in. NPTF tube fitting |
|---|
| Sample Streams | 1 |
|---|
| Sample Temperature | 5 to 45°C |
|---|
| Voltage | 85/132 VAC, 170/264 VAC |
|---|
| Weight (English) | 50 lb. |
|---|
| Weight (Metric) | 22.7 g |
|---|
| Width (English) | 17 in. |
|---|
| Width (Metric) | 45 cm |
|---|
| Temperature (Metric) Operating | 5°C to 35°C |
|---|
| Type | Sodium Analyzer |
|---|
| Unit Size | Each |
|---|