| Type | Plates and Racks |
|---|---|
| System Type | UPLC |
| Instrument | ACQUITY UPLC H-Class System |
| Replacement Schedule | Annual |
| Component | ACQUITY UPLC ISM |
| UNSPSC | 41115854 |
| Brand | ACQUITY, ACQUITY UPLC |
| Units per Package | 1 pk |
Answer:The I-Class Binary Solvent Manager (BSM).
Answer:Pump plungers, check valves, seals and filtration components.
Answer:No. The solvent manager and sample manager have separate preventive-maintenance kits.
The Waters ACQUITY I-Class Binary Solvent Manager performance maintenance kit combines routine service components for the UPLC binary pump module. It supports preventive maintenance of the solvent-delivery mechanism and includes pump plungers, check valves, high-pressure and wash seals, and filtration components. These items address wear and contamination in the solvent manager rather than the sample-injection path. The kit is intended for the I-Class BSM, not the I-Class fixed-loop or flow-through-needle Sample Manager, which use different maintenance consumables. It helps organize a planned pump service around matched Waters replacement parts.