Description
For Use With (Equipment): Thermo Scientific™ Evolution Pro UV-Visible Spectrophotometer
Temperature Range (Metric): −150 to +600°C
Unit Size: Each
Optional reaction chambers allow you to access temperatures from liquid nitrogen to 600°C and pressures from high vacuum to 500 psi. Run reaction gases through the system to study catalysts under reaction conditions. High-Performance Design
• Uses all-aluminum coated optics rather than traditional integrating spheres in its design
• Integrating spheres limit the wavelength range in the ultraviolet region
• All-specular optical system uses the full wavelength range of the instrument down to 190nm.
• Can be used on the Evolution 300, which has an upper wavelength range of 1100nm, providing extended wavelength coverage in the NIR region without having to purchase a more expensive UV-Visible or NIR instrument Flexible Sampling Options
• The sample chamber can be purged with an inert gas such as dry nitrogen, which allows the characterization of samples that may be oxygen- or moisture-sensitive
• Optional reaction chambers are also available for the study of samples at a wide variety of temperatures and pressures
• Temperatures can be controlled from -150° to +600°C, and variable vacuum control is available Convenient Analysis
• Can be used with the sample compartment open for added convenience when mounting and removing samples