| For Use With (Equipment) | 3500xL Genetic Analyzer, 3130<i>xl</i> Genetic Analyzer, 3130 Genetic Analyzer, 3500 Genetic Analyzer |
|---|---|
| Label or Dye | 6-FAM, LIZ, NED, SID, TAZ, VIC |
| Product Type | DS-36 Matrix Standard |
| Quantity | 8 Runs |
| Shipping Condition | Wet Ice |
| Software Type | Data Collection Software |
| Starting Material | DNA |
| Unit Size | Each |
Answer:The standard contains fragments labeled with 6-FAM, VIC, NED, SID, TAZ, and LIZ dyes.
Answer:It is used for spectral calibration so compatible genetic analyzers can distinguish the six fluorescent dye channels.
Answer:It should be stored refrigerated at 2–8°C.
DS-36 Matrix Standard (Dye Set J6) is used to perform spectral calibration on compatible capillary electrophoresis genetic analyzers before analysis of DNA fragments labeled with six fluorescent dyes. The standard contains six DNA fragments, each carrying a different label from the 6-FAM, VIC, NED, SID, TAZ, and LIZ dye set. During calibration, the instrument uses these defined signals to build the multicomponent matrix needed to distinguish overlapping fluorescence from the six dye channels in a single capillary. This calibration supports fragment analysis workflows on compatible SeqStudio, 3130, 3500, and 3730 series systems. The package provides material for approximately eight runs and is stored at 2–8°C. Matrix calibration is performed when required by the instrument workflow rather than for every sample injection, helping maintain reliable color separation for multiplex fluorescent fragment analysis.
Unless otherwise stated, the minimum order amount is USD 1,000.
For orders below USD 1,000, an additional handling fee may apply, or we may arrange local delivery / local handover where available.
All quotations are subject to final confirmation, including product availability, lead time, batch information, shipping cost, and applicable export or handling requirements.
For quotation requests, please send an RFQ from website or contact:
Please send inquiries directly to us in the “To” field. Emails sent via BCC, mass mailing, or with unrelated recipients may be automatically filtered and will not be processed.