Thermo Fisher, A41046, 3730xl DNA Analyzer (latest version)
Quantity: 1 analyzer, Unit Size: Each
Catalog Number: A41046
Send an RFQ.
Category
| Dimensions | 100 cm (W) x 65 cm (D) x 93 cm (H) |
|---|---|
| Duration | 1 Year Limited Warranty on Parts and Labor |
| External Computer | Computer and Flat Panel Monitor |
| Format | 384-well plate, 96-well plate |
| High-throughput Compatibility | High-throughput Compatible |
| Memory | 16GB (2x8GB) 1600MHz DDR3 Non-ECC |
| Operating Environment | Humidity: 20%-80% (non-condensing), Temperature: 18°C-30°C, Humidity: 20 to 80% (non-condensing) |
| Operating System | Microsoft Windows 10 |
| Optics | In-capillary detection by dual-side illumination, Backside-thinned CCD for maximum optical sensitivity, Multiple Filter Sets: Z, E, G5, G5-RCT, Any4Dyeset, Any5Dyeset and S (optional) |
| Performance | XLRseq: 50cm - 900 QV20 bases/3 hours, RapidSeq: 36cm - 550 QV20 bases/36 min, GeneMapper™ Software: 36cm - up to 500 bp res. w/ 0.15 bp sizing precision in 34 min, FastSeq: 50cm - 700 QV20 bases/1 hour, StdSeq: 36cm - 700 QV20 bases/1 hour, LongSeq: 50cm - 850 QV20 bases/2 hours |
| Plate Handling | Plate Seal: Automatic on-board piercing (septa or polypropylene heat seal), Internal bar code reader, Integrated auto sampler and plate stacker |
| Polymer |
Unless otherwise stated, the minimum order amount is USD 1,000.
For orders below USD 1,000, an additional handling fee may apply, or we may arrange local delivery / local handover where available.
All quotations are subject to final confirmation, including product availability, lead time, batch information, shipping cost, and applicable export or handling requirements.
For quotation requests, please send an RFQ from website or contact:
Please send inquiries directly to us in the “To” field. Emails sent via BCC, mass mailing, or with unrelated recipients may be automatically filtered and will not be processed.
| POP-7, POP Conformational Analysis Polymer |
| Polymer Consumption | 36 cm: 200μl⁄run, 50 cm: 250μl⁄run |
|---|
| Power Requirements | Voltage: 200-220 V or 230-240 V ±10%, Power Dissipated: About 2000 watts, Current: 15 amps max. Requires 30 amp receptacle w/in 3 m (10 ft) of instrument back. |
|---|
| Quantity | 1 analyzer |
|---|
| Reagent Handling | On-board reagents for up to 48 hrs unattended operation (with certain run modules), Automated polymer replenishment pump for increased throughput, Run modules: 5 sequencing, 1 fragment analysis and optional SNPlex™ |
|---|
| Sample Volume | 10-50 μL for 96-well plate, 5-30 μL for 384-well plate |
|---|
| Shipping Condition | Room Temperature |
|---|
| Temperature Range (Metric) | Temperature-controlled oven (18-70°C range) and detection cell heater |
|---|
| Weight | 180 kg (computer - 30 kg) |
|---|
| Capacity | 16 Sample Plates (96 or 384-well) |
|---|
| Capillary Length | 36 or 50 cm |
|---|
| For Use With (Equipment) | Real-Time PCR Systems or Thermal Cyclers, Sequencing Data Analysis |
|---|
| Number of Capillaries | 96 Capillaries |
|---|
| Product Line | Applied Biosystems |
|---|
| Read Length | 400 to 900 Bases |
|---|
| Throughput | 96 |
|---|
| Type | DNA Analyzer |
|---|
| Unit Size | Each |
|---|