Description
Measurement Modes: Temperature
Unit Size: Each
Features
• Large screen in a compact footprint
• meter accepts pH and Redox electrodes, in addition to Conductivity/TDS – it's like having fourmeters in one!
• Up to 5-point push button calibration
• Selectable cell constant
• Auto-ranging across 5 Conductivity ranges
• Easy interchangeability of probes
• Non-volatilememory holds up to 100 data points
• Integral electrode holderApplications
• Water quality testing
• Checkingmetal finishes
• Solutions
• Cooling tower water
• Printing fountain solutions
• Boiler water
• Brines
• Drillingmud
• Rinse tanks
• Ponds
• Pollution control
• Recirculating systems
• Waste water
• Industrial process systems
• General use in laboratories
• Titration
• Quality assurance
• Ecological studies
• Food-processingSpecificationsProduct Specification: DescriptionpH
• Range: -2.00 to 16.00 pH
• Resolution: 0.01 pH
• Accuracy: ±0.01 pH
• Cal. Points: Up to 5
• Buffer Sets: USA (1.68, 4.01, 7.00, 10.01, 12.45), NIST (1.68, 4.01, 6.86, 9.18, 12.45)ORP
• Range: ±2000mV
• Rel.mV Range: ±2000mV
• Resolution: 0.1mV (±199.9mV) / 1mV (beyond)
• Accuracy: ±0.2mV (±199.9mV) / ±2mV (beyond)Conductivity
• Range: 0.0 μS to 200.0mS
• Resolution: 0.01 / 0.1 / 1 μS ; 0.01 / 0.1mS
• Accuracy: ±1 % full scale
• Cal. Points: Automatic (84 μS, 1413 μS, 12.88mS, 111.8mS);maximum 1 per range;manual (5 points);maximum 1 per range
• Cell Constant: 0.1, 1.0, 10.0 (selectable)TDS
• Range: .. to 100.0 ppt @ 0.5 fact (200.0 ppt @ 1 factor)
• Resolution: 0.01 / 0.1 / 1 ppm ; 0.01 / 0.1 ppt
• Accuracy: ±1 % full scale
• Cal. Points:manual (5 points);maximum 1 per range
• TDS Factor: 0.40 to 1.00 (adjustable)Temperature
• Range (meter): 0.0 to 100.0 °C / 32.0 to 212.0 °F
• Resolution: 0.1 °C / 0.1 °F
• Accuracy: ±0.5 °C / ±0.9 °F
• Coefficient (Per °C): 0.00 to 10.00 %
• Normalization: 15.0 to 30.0 °C (adjustable)
• Compensation Type & Range: Automatic with supplied cell ormanual, 0.0 to 100 °C / 32.0 to 212.0 °F, 0 to 80 °C /, 32.0 to 176.0 °F with supplied cell
• Power Requirements: 9 V DC adapter, 1.3 A (100/240 VAC, SmPS)