| Description | Vanquish Diode Array Detector FG |
|---|---|
| Detection Mode | UV-vis Detection (Up to 10 Channels) |
| Drift | < 1 mAU/hour at 254 nm |
| Electrical Requirements | 100–240 V AC, 50/60 Hz, max. 245 W / 255 VA |
| Flow Cell Volume | Standard, path length 10 mm (13 μL, SST)<br /> Standard, biocompatible, path length 10 mm (13 μL, PEEK)<br /> Semi-analytical, path length 7 mm (5 μL, SST)<br /> Semi-micro, pathlength 7 mm (2.5 μL, SST)<br /> Semi-micro, biocompatible, path length 7 mm (2.5 μL, titanium)<br /> Diagnostic cell |
| GLP Features | Predictive performance functions: lamp age and ignitions, lamp intensity degradation, leak detection, service monitoring period.<br /> All system parameters logged in the Chromeleon CDS Audit Trail. |
| Lamp | Deuterium lamp, tungsten lamp |
| Linearity | < 5% at 2.2 AU (typically < 5% at 2.7 AU) |
| Max. Cell Operating Pressure | Standard biocompatible: 5 MPa (50 bar, 720 psi)<br /> Other flow cells: 12 MPa (120 bar, 1740 psi) |
| Noise | <± 6 μAU at 254 nm (standard flow cell, bandwidth: 4 nm, slit width: wide, time constant: 2 sec) |
| No. of Channels | 10 + 3D field |
Answer:It is a Vanquish diode array detector, not a multiple-wavelength detector.
Answer:Its wavelength range is 190–800 nm.
Answer:The detector supports up to ten absorbance channels and spectral acquisition.
Vanquish™ Diode Array Detectors and Multiple Wavelength Detectors is designed for UV-visible detection in Vanquish liquid chromatography and ultrahigh-performance liquid chromatography systems. The exact VF-series configuration is a diode array detector with a wavelength range of 190–800 nm; it is not the multiple-wavelength detector or a column compartment. Diode-array acquisition enables chromatographic peaks to be examined at several wavelengths while also collecting spectral information, supporting analyte identification and monitoring of minor impurities beside higher-concentration main components. The detector supports up to ten absorbance channels and a full spectral field during acquisition, with data rates up to 250 Hz under supported conditions. Optical design and low-noise electronics help limit baseline drift, and integrated wavelength verification uses a holmium-oxide filter. Baseline drift is less than 1 mAU per hour at 254 nm. Available flow-cell formats include a standard stainless-steel cell with a 10 mm path length and 13 µL volume and a biocompatible PEEK alternative, together with semi-micro configurations. This detector provides spectral and absorbance measurements within an assembled LC system; it does not independently pump samples or perform chromatographic separation. Spectral data can complement single-wavelength chromatograms when overlapping peaks or impurity signals require additional absorbance information. A diode-array design therefore provides more information than a detector limited to one fixed wavelength.
Unless otherwise stated, the minimum order amount is USD 1,000.
For orders below USD 1,000, an additional handling fee may apply, or we may arrange local delivery / local handover where available.
All quotations are subject to final confirmation, including product availability, lead time, batch information, shipping cost, and applicable export or handling requirements.
For quotation requests, please send an RFQ from website or contact:
Please send inquiries directly to us in the “To” field. Emails sent via BCC, mass mailing, or with unrelated recipients may be automatically filtered and will not be processed.
| PC Connection |
|---|
| USB 2.0; 3-port HUB to connect further Vanquish modules |
| Photodiode Array | 1024 element |
|---|
| Safety Features | Power-up diagnostics of optics, cooling fans, motors and electronics. Leak detection and safe leak handling. |
|---|
| Slit Width | Settable: Wide, narrow |
|---|
| Spectral Bandwidth | Pixel resolution: 0.6 nm (average) |
|---|
| Temperature Range | 5 to 35°C; 20 to 80% RH (non-condensing) |
|---|
| Validation Functions | Internal validation with holmium oxide filter |
|---|
| Wavelength Accuracy | ± 1 nm |
|---|
| Wavelength Range | 190–800 nm |
|---|
| Wavelength Repeatability | ± 0.1 nm |
|---|
| Weight (Metric) | 16.3 kg |
|---|
| Data Collection Rate | 250 Hz Max. |
|---|
| Detector Type | Single-Beam |
|---|
| Type | Diode Array Detector |
|---|
| Unit Size | Each |
|---|