Description
General description
Determined at 10% breakthrough by frontal analysis at a mobile phase velocity of 500 cm/h in a column with a bed height of 20 cm.
HiTrap® MabSelect™ Xtra columns are prepacked with MabSelect™ Xtra, a BioProcess medium for capture of monoclonal antbodies from large sample volumes.
pH below 3 is sometimes required to elute strongly bound IgG species. However, protein ligands may hydrolyze at very low pH.
Application
The increased capacity of the medium is well suited for purification of Mabs from high-level expression feedstock.
Features and Benefits
Increased dynamic binding capacity with smaller particle size and greater porosity than MabSelect™ medium.
MabSelect™ Xtra uses the same recombinant Protein A ligand as MabSelect™
Oriented coupling of recombinant Protein A to the matrix via an engineered C-terminal cysteine enhances IgG binding capacity.
Fewer regulatory concerns due to the total absence of mammalian culture in the ligand production and purification.
Prepacked HiTrap® columns for process development, screening of purification conditions, and small-scale purification of Mabs.
Preparation Note
Please be aware this product may be shipped 90 days before the expiration date. For more information on the batch specific expiration date, please contact technical service.
Analysis Note
To view the Certificate of Analysis for this product, please visit www.cytiva.com.
Other Notes
The pressure over the packed bed varies depending on a range of parameters such as the characteristics of the chromatography medium and the column tubing used.
Legal Information
HiTrap is a registered trademark of Cytiva
MabSelect is a trademark of Cytiva